Tech Specs
Defibrillator
TYPE
Semi-automatic external
defibrillatorMODELS
DDU-100A, DDU-100EWAVEFORM
Biphasic Truncated Exponential
(Impedance compensated)ENERGY
Adult: 150 Joules
Pediatric: 50 Joules
(nominal into 50 ohm load)CHARGE TIME**
DBP-2800: Less than 6 seconds
DBP-1400: Less than 9 seconds
**Typical, with new battery at 25º CVOICE PROMPTS
Extensive voice prompts
guide user through operation
of the unit and CPR
CPR PACING
MetronomeCONTROLS
– Lighted On/Off button
– Lighted Shock buttonINDICATORS
– “check pads”
– “do not touch patient”
– “analyzing”
– AED status LED
Self TestsAUTOMATIC
Automatic daily, weekly and
monthly and quarterly circuitry testsBATTERY INSERTION
System integrity test on battery
insertionPAD PRESENCE
Pads preconnected tested dailyUSER-INITIATED
Unit and battery pack system
test may also be initiated by the
userSTATUS INDICATION
Visual and audible indication of
unit statusPatient Analysis System
PATIENT ANALYSIS
Automatically evaluates patient impedance for proper pad contact. Monitors signal quality and analyzes patient ECG for shockable/non-shockable rhythmsSENSITIVITY/SPECIFICITY
Meets AAMI-DF-39 specifications and AHA recommendations
Defibrillation / Monitoring PadsMODEL
Adult – DDP-100
Child/Infant – DDP-200PTYPE
Pre-connected, single-use, non-polarized, disposable, self-adhesive electrodes with cable and connectorSURFACE AREA
103 cm2 (nominal, each pad)
50 cm2 (nominal, each pad)PAD PLACEMENT
Adult – Anterior/Anterior
Child/Infant – Anterior/PosteriorCABLE LENGTH (typical)
48 in (122 cm)Environmental
TEMPERATURE
Operating: 0 to 50°C (32 to 122°F)
Standby: 0 to 50°C (32 to 122°F)
(with battery installed)RELATIVE HUMIDITY
Operating / Standby: 5% – 95%
(non-condensing)ALTITUDE
-500 to 15,000 ft (-150 to 4500 m) per MIL-STD-810F 500.4 Procedure IIVIBRATION
Ground (MIL-STD-810F 514.5 Category 20)
Helicopter (RTCA/DO-160D, Section 8.8.2, Cat R, Zone 2, Curve G)
Jet Aircraft (RTCA/DO-160D, Section 8, Cat H, Zone 2, Curves B & R)SHOCK / DROP ABUSE TOLERANCE
MIL-STD-810F 516.5 Procedure IV (1 meter, any edge, corner, or surface, in standby mode)SEALING / WATER RESISTANCE
IEC60529 class IP54; Splash Proof, Dust Protected (Battery Pack installed)ESD
EN61000-4-2:2009 (open air up to 15kV or direct contact up to 8kV)EMC (Emission)
EN60601-1-2:2001 +A1:2006, limits EN55011:1998 +A1:1999 +A2:2002 Group 1 Level B, methodEMC (Immunity)
EN60601-2-4:2003, limits EN 61000-4-3:2002 Level 3 (10V/m), methodBattery Pack
TYPE
– Lithium/Manganese Dioxide
– Disposable, recyclable, non rechargeableLOW BATTERY INDICATORS
– Visible
– AudibleMODEL DBP-2800
POWER
15V, 2800 mAhCAPACITY **
– 300 shocks or 16 hours continuous operationSTANDBY-LIFE**
– 7 yearsMODEL DBP-1400
POWER
15V, 1400 mAhCAPACITY**
– 125 shocks or 8 hours
continuous operationSTANDBY-LIFE**
– 5 years**Typical, with new battery at 25º C
Event DocumentationINTERNAL EVENT RECORD
Critical ECG segments and rescue event parameters are recorded and can be downloaded to a removable data cardPC-BASED EVENT REVIEW
ECG with event tag display, and audio playback when availableREMOVABLE STORAGE
(optional) Up to 12 hours of ECG and event data storage (no audio option) or up to 2 hours of audio, ECG and event storage (audio option) on a removable data card. Actual length of storage is dependent on card capacity.
SIZE
8.5 x 11.8 x 2.7 inches
(22 x 30 x 7 cm)WEIGHT (Approximate)
With DBP-1400: 4.2 lbs (1.9 kg)
With DBP-2800: 4.4 lbs (2.0 kg)